Responding to a Potassium Deficiency on the CMS-2567
A strong plan of correction for a Potassium finding names the fix, shows a look-back for affected patient results, changes the system that failed, assigns monitoring and gives a completion date, with the Potassium records that prove it attached.
What a Potassium plan of correction should cover
CMS returns plans that only promise to do better. Plans of correction generally need to explain:
- The fix: what was corrected for Potassium and when.
- The look-back: how you found other Potassium results or patients that may have been affected, and what you did about them.
- The system change: the procedure, training or software change that keeps it from happening again.
- The monitor: who checks it, how often, and where it is reported (for example to the laboratory director or QA meeting).
- The date: a realistic completion date for each item.
Evidence to attach for Potassium
- Daily Potassium QC records, Levey-Jennings charts, corrective action log, control lot ranges
- Six-month Potassium calibration verification and reportable range data with director review
- Potassium PT enrollment, attestation statements, graded reports, investigation of every miss
- Potassium verification study: accuracy, precision, reportable range, reference interval check, director sign-off
- Critical value list, call log with read-back, report format showing Potassium reference intervals
- Training and competency records for everyone who runs Potassium
Potassium PT and accuracy
Potassium is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: target value ±0.3 mmol/L. If the finding involves PT, include the investigation of each unacceptable result and the remedial action taken; routine chemistry rules require that documentation be kept for two years (42 CFR 493.841).
Common root causes behind Potassium findings
- Hemolysis (in vitro) during collection or transport releases intracellular potassium and falsely raises results; use a hemolysis index and a reject or comment rule.
- Fist clenching and prolonged tourniquet time during venipuncture raise potassium.
- Delayed separation, and refrigeration of unspun whole blood, raise potassium; centrifuge within your defined time.
- Procedures that do not match what staff actually do.
- No second-level review of QC and calibration verification records.
Related tests
- Sodium (full guide)
- Chloride
- Carbon Dioxide (CO2, Bicarbonate) (full guide)
- Creatinine
- Magnesium (full guide)
Full guide: Potassium testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Received a 2567? Talk it through with a CLIA consultant
Sound Lab Consultants reviews CMS-2567 findings with laboratory directors, works through root causes and helps draft plans of correction and allegations of compliance.
Related guides on soundlabconsult.com:
Frequently asked questions
How do I write a plan of correction for a Potassium deficiency?
State the correction, the look-back for affected Potassium results, the system change, the monitoring and a completion date, and attach the supporting records.
What records support a Potassium plan of correction?
Daily Potassium QC records, Levey-Jennings charts, corrective action log, control lot ranges; Six-month Potassium calibration verification and reportable range data with director review; Potassium PT enrollment, attestation statements, graded reports, investigation of every miss.
Where is the plan of correction written?
On Form CMS-2567, the Statement of Deficiencies and Plan of Correction, or in the format your state agency or accrediting organization specifies.
Sources
- 42 CFR 493.931, Routine chemistry
- Form CMS-2567, Statement of Deficiencies and Plan of Correction
- CMS State Operations Manual, Chapter 6 (Special Procedures for Laboratories)
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493 Subpart H, Participation in Proficiency Testing
- CMS State Operations Manual, Appendix C (Laboratories)