Responding to a Cystatin C Deficiency on the CMS-2567
A strong plan of correction for a Cystatin C finding names the fix, shows a look-back for affected patient results, changes the system that failed, assigns monitoring and gives a completion date, with the Cystatin C records that prove it attached.
What a Cystatin C plan of correction should cover
CMS returns plans that only promise to do better. Plans of correction generally need to explain:
- The fix: what was corrected for Cystatin C and when.
- The look-back: how you found other Cystatin C results or patients that may have been affected, and what you did about them.
- The system change: the procedure, training or software change that keeps it from happening again.
- The monitor: who checks it, how often, and where it is reported (for example to the laboratory director or QA meeting).
- The date: a realistic completion date for each item.
Evidence to attach for Cystatin C
- Daily Cystatin C QC records, Levey-Jennings charts, corrective action log, control lot ranges
- Six-month Cystatin C calibration verification and reportable range data with director review
- Twice-yearly Cystatin C accuracy checks (external PT, split samples or reference material) with acceptance criteria
- Cystatin C verification study: accuracy, precision, reportable range, reference interval check, director sign-off
- Critical value list, call log with read-back, report format showing Cystatin C reference intervals
- Training and competency records for everyone who runs Cystatin C
Cystatin C PT and accuracy
Not listed in Subpart I: no required PT, but accuracy must be verified at least twice a year (42 CFR 493.1236(c)). If the finding involves PT, include the investigation of each unacceptable result and the remedial action taken; routine chemistry rules require that documentation be kept for two years (42 CFR 493.841).
Common root causes behind Cystatin C findings
- Calibrator traceability to the international reference material matters for eGFR equations.
- Verify reportable range and reference intervals for your method.
- Procedures that do not match what staff actually do.
- No second-level review of QC and calibration verification records.
Related tests
Full guide: Cystatin C testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Received a 2567? Talk it through with a CLIA consultant
Sound Lab Consultants reviews CMS-2567 findings with laboratory directors, works through root causes and helps draft plans of correction and allegations of compliance.
Related guides on soundlabconsult.com:
Frequently asked questions
How do I write a plan of correction for a Cystatin C deficiency?
State the correction, the look-back for affected Cystatin C results, the system change, the monitoring and a completion date, and attach the supporting records.
What records support a Cystatin C plan of correction?
Daily Cystatin C QC records, Levey-Jennings charts, corrective action log, control lot ranges; Six-month Cystatin C calibration verification and reportable range data with director review; Twice-yearly Cystatin C accuracy checks (external PT, split samples or reference material) with acceptance criteria.
Where is the plan of correction written?
On Form CMS-2567, the Statement of Deficiencies and Plan of Correction, or in the format your state agency or accrediting organization specifies.
Sources
- Form CMS-2567, Statement of Deficiencies and Plan of Correction
- CMS State Operations Manual, Chapter 6 (Special Procedures for Laboratories)
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493 Subpart H, Participation in Proficiency Testing
- CMS State Operations Manual, Appendix C (Laboratories)